Calibration

Mechanical, thermal and electro thermal calibration

Thermal

Quantity Measured / Instrument Range/Frequency *Calibration Measurement Capability (±) Remarks
Liquid in Glass Thermometer, Temperature Gauge

RTD/Thermo couple with or without indicator.
-80°C to 50°C 1.1 °C Using SSPRT/RTD indicator with Temperature Indicator & 6.5 DMM/Beamex & liquid & Dry bath circulators.

Using R-type thermocouple with Indicator & 6.5 DMM and Dry-Block Calibrator by Comparison Method.
50°C to 300°C 1.0 °C
300°C to 600°C 2.2 °C
600°C to 1200°C 3.5 °C
Internal Temperature Sensor with Indicator, Thermo hygrometer 15°C to 50°C 0.80°C Using Temperature Sensor with Indicator & Cole Parmar Incubator by Comparison Method
% RH Indicator with Sensor, Thermo hygrometer 22% RH to 95% RH @ 25°C 2.7% Using Fixed RH Salt solution with Digital Temp/RH Indicator with sensor, by comparison method.
Temp Indicator with sensor of Oven, Incubator, Muffle Furnace, Water Bath, Deep Freezer, COD, BOD, Refrigerator, RTD/Thermo couple with or without Indicator -80°C to 50°C 1.1 °C Using SSPRT/RTD indicator with Temperature Indicator & 6.5 DMM/ Beamex & Liquid bath circulators, by Comparison Method

Using R-Type Thermocouple with Indicator& 6.5 DMM and Dry-Block Calibrator by Comparison Method
50°C to 300°C 1.0 °C
300°C to 600°C 2.2 °C
600°C to 1200°C 2.8 °C
Humidity Chamber 22% RH to 95% RH @ 25°C 2.7% Using Digital Temp/RH Indicator with sensor, by comparison method.