Calibration
Mechanical, thermal and electro thermal calibration
Electro-Technical
Quantity Measured / Instrument | Range/Frequency | *Calibration Measurement Capability (±) | Remarks |
DC Voltage# | 1 mV to 100 mV | 0.81 % to 0.01 % | Using Fluke 6½ DMM 8846A with source 5500A MFC By Direct/ Comparison Method |
100 mV to 100 V | 0.01 % to 0.008 % | ||
100 V to 1000 V | 0.008 % to 0.01 % | ||
DC Current# | 100 μA to 100 mA | 0.1 to 0.22 % | |
100 mA to 10 A | 0.22% |
Resistance (4 wires & 2 wires)# | 1 Ω to 100 KΩ | 0.8 % to 0.016% | Using Fluke 6½ DMM 8846A / 5500A MFC By Direct Method |
100 KΩ to 10 MΩ | 0.016 % to 0.082% | ||
10 MΩ to 100 MΩ | 0.082 % to 1.0 % |
AC Voltage# | 50 Hz 100 mV to 1000 V | 0.122 % to 0.13 % | Using Fluke 6½ DMM 8846A with source 5500A MFC By Direct/ Comparison Method |
AC Current# | 50 Hz 100 μA to 1 A | 0.5 % to 0.2 % | Using Fluke 6½ DMM 8846A with source 5500A MFC By Direct/ Comparison Method |
1 A to 10 A | 0.2 % to 0.26 % |
Frequency# | 45 Hz to 100 kHz | 0.018 % to 0.058% | Using Fluke 6½ DMM 8846A with source 5500A MFC By Direct/ Comparison Method |
Capacitance# | 10nF to 1 mF | 6.0 % to 3.0 % | Using Fluke 6 ½ DMM 8846A / 5500A MFC, By Direct Method |
Temperature Simulation RTD Type# | (-) 200°C to 600°C | 0.67°C | Using Fluke 6 ½ DMM 8846A, By Direct Method |
Stop Watch / Timer / Digital Clock# | 40 s to 24 hours (LC: 0.01 s) | 0.57s to 5.845s | Using Digital Stop Watch by Comparison Method |